Situation:
You need to perform a DICOM calibration on your Eizo Radiforce medicaid display monitors.
Solution:
Follow the instructions below to calibrate your Eizo monitors using RadiCS v5. Please contact us to request a copy of the Eizo calibration and QA software if you need a copy. Please follow the instructions on how to install the Eizo RadiCS calibration software if you do not have a copy installed on your system.
Before you start, please make sure that your USB Type-B cable is connected from
your monitors to your computer.
Open RadiCS Version 5
Wait for RadiCS to detect monitor(s)
Click on Administrator mode
Type password: passwordv5. Click OK
Click Calibration
Verify the monitors are selected. Click Proceed
During the calibration process, the RadiCS software will fade out
Also, during the calibration, the monitor(s) will go black and the build-in sensor will pop out.
On the Finish page, click on Passed or Failed to view a detailed report
Click Finish
Conclusion:
Your monitor(s) will now be calibrated to the DICOM standard. If you wish to configure RadiCSv5 to automatically calibrate your monitors, please follow the How to Schedule Automatic Calibration in Eizo RadiCS DICOM Calibration and QA Software v5 - Win 10 Pro guide.